dc.contributor.author |
UMANINGRUM, DEWI |
|
dc.date.accessioned |
2021-11-22T01:15:14Z |
|
dc.date.available |
2021-11-22T01:15:14Z |
|
dc.date.issued |
2021-09-29 |
|
dc.identifier.uri |
https://repo-dosen.ulm.ac.id//handle/123456789/21931 |
|
dc.description.abstract |
Research on the Ion-Selective Electrode (ISE) of coated wire-type tartrazine using chitosan as
an ionophore has been developed. The variables used in the manufacture of ISE are membrane
composition and immersion time. Meanwhile, the basic characteristics of ISE measured are
Nernst value, measurement concentration range, detection limit, and measurement response
time. The results showed that ISE tartrazine coated wire type had an optimum membrane
composition in a mixture of chitosan: PVC: DOP of 3: 34: 63 (% w/w) and a membrane
immersion time 20 minutes. The basic characteristics of ISE produce a Nernst value of 20.976
mV/decade. The measurement concentration range is 1x10-7 – 1x10-2 M with a detection limit
of 2.749x10-7 M or 0.1469 ppm. The response time ranges from10-60 seconds, with an average
of 40 seconds. |
en_US |
dc.publisher |
Universitas Lambung Mangkurat |
en_US |
dc.subject |
Research Subject Categories::TECHNOLOGY::Chemical engineering |
en_US |
dc.title |
The Coated-Wire Ion-Selective Electrode (CWISE) of Tartrazine Using Chitosan as an Ionophore |
en_US |
dc.type |
Article |
en_US |